Disposable Platelet Segregator
A.Category£ºBlood Transfusion
B.Compisition£º
It consists of segregative cup and pipe system.
C.Application£º
It is used to collect aseptic blood plasma and platelet and re-transfuse the
separated blood cells to the blood donor.
D.Specification£º
E.Technical requirements£º
1,The latex tube in the system is soft, transparent, clear and springy. Through
it air bubbles and blood flow can be seen.
2,The filtering rate of blood filter and drug solution filter is not less than
80%.
3,The filtering rate of pressure transducer air filter is not less than 90%
(dry state).
4,The segregator is transparent and clean and the content can be seen through
the wall.
5,Friction or deformation will not happen in two hours at 4000rpm.
6,All connection parts are connected firmly and well sealed.
7,All protective caps will not fail off easily but can be dismantled easily.
They keep the internal parts sterile.
8,The filtering rates of leucocyte , red blood cells and platelet in blood plasma
are 99.9%, 95% and 90% after separation respectively.
9,Its chemical performance complies with the standard.
10,It is sterilized with ethylene oxide and the quantity of ethylene oxide residue
complies with the standard.
11,The product is sterile, has no pyrogen and does not induce hemolysis reaction,
skin allergic reaction, cytotoxicity or acute systemic toxicity.